VB-LF6200 integrated intelligent sulfur measuring instrument

Picture Name
Coal, electric power, chemical industry, geological survey, environmental protection, quarantine inspection, scientific research and colleges and other related industries and departments. Provide accurate total sulfur data for arbitration.

Classification: Elemental Analysis Series

Email: vbjd2010@126.com

VB-LF6200 integrated intelligent sulfur measuring instrument
+
  • VB-LF6200 integrated intelligent sulfur measuring instrument

Technical parameters/Technical Parameters

1. Test method: coulometric titration.

2. Sulfur measurement range:(0.01~40)%, sulfur resolution: 0.001%

3. Test time:(3~6)min/sample, sample weight: 50-80mg

4. Heating rate: 50-100 ℃ /min according to relevant national standards

5. Accuracy: in line with and better than the national standard

6. Working furnace temperature: 1150±5 ℃, 900±5 ℃, temperature control accuracy: 1 ℃

7. Voltage: 220±22VAC 50±1HZ Power:<4Kw

8. Weight: 50Kg

9. Host size: 365x505x685mm

 

Application/Apply Domain

Coal, electric power, chemical industry, geological survey, environmental protection, quarantine inspection, scientific research and colleges and other related industries and departments. Provide accurate total sulfur data for arbitration.

 

Technical features/Technical Features

automatic control

1. Furnace insulation material: Class A mullite is used, and the constant temperature zone is long and stable.

2. Interface mode: TCP network communication protocol is adopted to determine that the instrument communication has no error rate.

3. Temperature control method: PID algorithm is adopted to prevent overshoot and temperature rise, which is safe, accurate in temperature control and fast in temperature rise.

4, air supply components: the use of new piston air pump, small size, low noise, long life, and the use of double flow meter horizontal control technology to make the gas sulfur long-term stability without personnel intervention without computer monitoring.

5. Mixing component: frequency conversion technology is adopted to control the mixing so that the mixing rotor will not jump up and hit the pole piece.

6. Control mode: bit machine control. More effective control of the host operation is not caused by the computer unilateral reasons uncontrollable. Fault alarm self-protection.

7, analysis and testing system: can be connected to the company's most coal quality analysis equipment. Can automatically upload data, production reports. (Users can design reports according to their own needs), can be connected to the balance.

High reliability of accurate results

1. Adopt a variety of integral working methods.

2. A unique system is used to automatically analyze the standard data for systematic linear error correction and sulfur trioxide fusion.

Product Message

If you are interested in our products, please leave your email, we will contact you as soon as possible, thank you!

Submit Message