VBR-6000/D automatic calorimeter
Classification: Calorimeter series
Email: vbjd2010@126.com
Technical parameters/Technical Parameters
Temperature measurement range: 0-40 ℃
Test time: regular method about 16min
Resolution: 0.0001 ℃
Rapid method about 10min
Accuracy: Within the uncertainty of the standard sample.
Precision: RSD≤ 0.1%
Voltage:(220±22)VAC (50±1)HZ
Power: ≤ 500W
Host size: host size: 605 x 505 x 505mm
Standard configuration: VBR-6000/D host/Lenovo brand computer/laser printer
Apply/Apply Domain
The product is used in electricity, energy, oil, environmental protection. The high-precision calorific value measurement of various solid and liquid substances in food and scientific research departments is especially suitable for the measurement of low calorific value combustible substances (fly ash, gangue, cement raw meal).
Technical features/Technical Features
1. Adopt vertical (horizontal) structure, 45KG water weight.
2. Highly intelligent, using isolation barrel to demarcate the outer barrel and the inner barrel, and configuring chip refrigeration technology to ensure that the temperature of each sample test is consistent.
3. The unique overflow technology of double-fixed container is adopted to determine the water volume, which ensures that the water volume tested by each sample is fully consistent.
4, the use of chip refrigeration technology. The temperature rise caused by cooling each sample tested.
5. Connection mode: VB-CAN card, using shielding and protection circuit device lightning protection, anti-magnetic interference, anti-stimulation interference.
6. Using PT1000 temperature measuring probe, temperature linearity is better.
7, analysis and testing system: can be connected to the company's most coal quality analysis equipment. Can automatically upload data, production reports. (Users can easily write according to their own needs), can be connected to the balance.
8, the use of power optimization circuit, so that the equipment works under the fixed frequency and pressure of sulfur.
9. Air heat transfer in the inner barrel, air heat transfer is better than water heat transfer, and internal circulating air technology is adopted. Accurate and reliable data
2, Using a unique system to automatically analyze the standard data for system linear error